Product Details:
Payment & Shipping Terms:
|
| Product Name: | Best Price Square Type UV Accelerated Weathering Chamber ASTMG53-77 UV Light Test | UV Temperature: | RT+10℃~70℃ |
|---|---|---|---|
| Temperature Uniformity: | ±3℃ Temperature Fluctuation: ±0.5℃ | Humidity Range: | ≥75%RH |
| Center Distance In Lamp Tube: | 65~70mm | Ultraviolet Tube: | UVA-340 Wavelength Range 290-400nm (its Peak Value Is 340nm) UVB-313 Wavelength Range 280-320nm (Its Peak Is 313nm) |
| UVA-340 Wavelength Range 290-400nm (its Peak Value Is 340nm) UVB-313 Wavelength Range 280-320nm (Its: | ASTM G 154, ISO 4892-1, ISO 11507, ASTMD 4329 GB/T 14522-93 | ||
| Highlight: | Square Type Accelerated Weathering Chamber,UV Light Accelerated Weathering Chamber,ASTMG53-77 UV Light Test Chamber |
||
| Parameter | Specification |
|---|---|
| Sample Mounting | Test samples are fixed on the test rack facing fluorescent lamps. Empty positions are filled with blackboards to maintain chamber closure. |
| Test Temperatures | Illumination: 50°C, 60°C, or 70°C (60°C preferred); Condensation: 50°C; Temperature uniformity: ±2°C |
| Cycle Options | 4 hours illumination / 4 hours condensation or 8 hours illumination / 4 hours condensation |
| Lamp Replacement | Replace fluorescent lamps after 400-450 hours of operation. Remaining lamps compensate for aging effects, ensuring 1600-1800 hours total service life. |
| Maintenance | When replacing lamps, clean and dry the water pan to prevent scale formation. |
Contact Person: Ms. Fiona Zhong
Tel: +86 135 3248 7540
Fax: 86-0769-3365-7986