Product Details:
Payment & Shipping Terms:
|
Product Name: | Lab Vacuum High Pressure Accelerated Aging Test Chamber Environmental Simulation Chamber | Setting Temperature: | +100 ℃~+132 ℃ (saturated Steam Temperature) |
---|---|---|---|
Humidity Range: | 100 % Saturated Vapor Humidity | Working Pressure: | 0.5 Kg ~ 2.5 Kg / Cm2 (the Inner Barrel Is Designed To Withstand 4 Kg / Cm2 ) |
Ime Range: | 000 Hr ~ 999 Hr | Pressurization Time: | 0.00 Kg ~ 1.04 Kg / Cm2 Within About 45 Minutes |
Test Box Size: | ∮340 Mm X L500 Is Mm, Round Test Box (can Be Customized) | Overall Dimensions: | 900x 900 X 1500 Mm ( W * D * H ) |
High Light: | SUS316 Accelerated Aging Test Chamber,High Pressure Accelerated Aging Test Chamber,SUS304 Environmental Simulation Chamber |
Lab Vacuum High Pressure Accelerated Aging Test Chamber Environmental Simulation Chamber
1.Performance:
1.1 Setting Temperature | +100 ℃~+132 ℃ (saturated steam temperature) |
1.2 Humidity Range | 100 % saturated vapor humidity |
1.3 Working Pressure | 0.5 Kg ~ 2.5 Kg / cm2 (the inner barrel is designed to withstand 4 Kg / cm2 ) |
1.4 Time Range | 000 Hr ~ 999 Hr |
1.5 Pressurization Time | 0.00 Kg ~ 1.04 Kg / cm2 within about 45 minutes |
2.Structure:
2.1 Test Box Size: ∮340 mm x L500 is mm, round test box (can be customized)
2.2 Overall Dimensions: 900x 900 x 1500 mm ( W * D * H )
2.3 Inner Barrel Material: stainless steel plate material # SUS-316
2.4 Outer Barrel Material: stainless steel plate material # SUS-304
2.5 Control System:
a. Using Japanese-made YOKOGAWA UT-150 microcomputer to control saturated steam temperature
b. The time controller adopts LED display.
c. Use the pointer to display the pressure gauge.
d. Microcomputer P.I.D automatic calculation to control the saturated steam temperature.
e. Manual water inlet valve (automatic replenishment function, continuous test without stopping)
2.6 Mechanical Structure:
a. Round inner box, stainless steel round test inner box structure, in line with industrial safety container standards, can prevent dew condensation and dripping design during the test.
b. Circular lining, stainless steel circular lining design, can avoid steam latent heat directly impacting the test sample.
c. Precise design, good air-tightness, low water consumption, continuous operation of 400Hrs each time adding water.
d.The patented packing design makes the door and the box more closely combined, which is completely different from the traditional extrusion type, which can prolong the packing life.
e. Critical point LIMIT mode automatic safety protection, abnormal cause and fault indicator display
3. Other Accessories:
3.1 1 set of test racks
3.2 Sample Tray
4. Power Supply System:
4.1 The fluctuation of the system power supply shall not be greater than ±10
4.2 Power supply: single phase 220V 20A 50/60Hz
5. Environment & Facilities:
5.1 The allowable working environment temperature is 5℃~30℃
5.2 Experimental water: pure water or distilled water
6. Attached Documents:
6.1 System brand new warranty, warranty
6.2 Operation and Maintenance Manual
Contact Person: Ms. Fiona Zhong
Tel: +86 135 3248 7540
Fax: 86-0769-3365-7986